1.Description
Non-contact Body Infrared Thermometer is Specially designed fo take the body temperature of a person regardless of room temperature
Depending on various skin types and thickness, there may be temperature difference When the indoor temperature is significant difference, the body infrared thermometer should be placed in that room for 15 to 20 minutes before using.
2.Safety Information
•This device must only be used for the purposes described in this instruction manual.
•This device must only be used in an ambient temperature range between 10°C and 40°C.
•Do not expose this thermometer to extreme temperature conditions of>40 C or<0 C.
•Do not expose this thermometer to electric shocks
•Do not use the device in relative humidity higher than 85%RH
•Do not use the device near large electromagnetic fields such as found with cordless or cell phones.
•Keep the device away from water and heat, including direct sunlight
•Do not drop or knock the device, and dot not use if damaged
•It may affect the accuracy of measurements when the forehead is covered by hair perspiration, cap or scarf
•Keep the measuring distance as <5cm(<2in)
•When the body infrared thermometer should be left in that room during 15to 20 minutes before using.
•It may affect the accuracy of measurements when the forehead is covered by perspiration
or other factors, please take the temperature behind the ear lobe.
•Clean the glass with cotton bud lightly moistened with 70% alcohol
3.Features
•Precise non-contact measurements
•Built-in laser pointer
•User selectable °C or °F(Th rough the” MO DE” key long press mroe than 5 sec)
•Automatic selection range and Display Resolution 0.1°C(32.1°F)
•Autondsomatic Data Hold & Auto Power off
•Set Alarm value(the alarm threshold default value is 38°C(100.°F)
Instruction: turn on unit and pressnMODEHbutton 3 seconds, LCD display”Fl
Re-press”MODE”to enter 38°C(dufaIt alarm setting), increase or decrease value
By press up ordown arrow, press measurement rigger to finish setting.
•Selection for buzze
Instruction: turn on the unit and press”MODE”button 3 seconds, LCD display”F3
Re-press” MODEHfor setting, press up ordown arrow for selection,,ON”or,,OFF”mode. press measurement rigger to finish setting
•Selection for body or ambient temperature
Operation Way: In Standby, Press the” Mode button, then the LCD screen showing”Fl 38.0°C,r, by pressing the” <to>”to change the Data, Enter into F2 setting by press MODE”one more time, then the LCD screen showing”F2 0.0°C”, you can press,toA” to change the Data, It is the same operation way for F3 and F4 Note: “5URF”For showing the object, “Body for the human being’s brow mode under the F4 setting
4.Intended Use
•The Non-contact Body Infrared Thermometer is designed for body surface and forehead temperature measurement for infants and adults without contacting to human body. It can be used by consumers in household environment in accordance with the ASTM 1965-1988
•Standard specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
5.Precaution Before Use
•In order to obtain reliable and stable results, when changing the environment to a room with a significant temperature difference, it is recommended to leave the Non-contact Boy Infrared Thermometer to this ambient temperature for 15 to 20 minutes before using it
•Depending on various skin types and thickness, there may be temperature difference.
•Do not expose the thermometer to sunlight or to water.
Allow one minute interval between two measurement.
Normal Temperatures According To Measurement Method
Measurement |
Methodnormal Temp °C |
Normal Temp°F |
Rectal |
36.6 to 38 |
97.8 to 100.4 |
Oral |
35.5 to 37.5 |
95.9 to 99.5 |
Auxiliary |
34.7 to 37.3 |
94.4 to 99.1 |
Auriclar |
35.8 to 38 |
96.4 to 100.4 |
Temporal(Thermo Flash) |
35.8 to 37.8 |
96.4 to 100.0 |
Age |
Temp °C |
Temp°F |
0-2 years | 34.6 to 38 | 97.8 to 100.4 |
3-10 years |
36.1 to 37.8 |
97.0 to 100.0 |
11-65 years |
35.9 to 37.6 |
96.6 to 99.7 |
>65 years |
35.8 to 37.5 |
96.4 to 99.5 |
7. Technical Specifications
Normal Conditions of Use:
Display Resolution | 0.1oC(32.1oF) |
Operating Temperature | 10to40°C(50tol04.°F) |
Humidity Rate | ≤85% |
Power | R03AAA1.5V2 |
Size | 95×42.5xl56mm(LXWXH) |
Weight | 124g |
Measuring Range:
Body Measuring Range |
32.0to43.0°C(90tol09.4°F) |
Ambient Measuring Range | Oto 60°C(32tol40°F) |
Body Accuracy | ±0.3°C(32.54°F) |
Ambient Accuracy | ±1°C(33.8°F) |
Measuring Distance |
<5cm (2 in) |
Automatic Stop |
7 sec |
Non-contact Body Infrared Thermometer Precision:
32to35.9°C/93.2to96.6°F | ±0.3°C/32.9°F | According to ASTMStandard E1965-1998(2003) |
36to 39°C/96.8to 102.2°F | ±0.2°C/32.36°F | |
39 to 42.5/102.2 to 108.5°F | ±0.3°C/32.9°F |